CVE-2026-18844
Pulsetto Vagus Nerve Stimulator Hidden Functionality
CVSS 4.0 7.2 HIGHCVSS 3.1 8.1 HIGHCheck source for fixed versionsKEV status unknown
cve-program · CVE-2026-18844
PublishedThe firmware of the Pulsetto Vagus Nerve Stimulator accepts several undisclosed commands over its Bluetooth Low Energy (BLE) interface. These commands are sent without authentication or encryption, and are never issued by the companion mobile application, yet are fully processed by the device when it is powered on.
Read the publisher's advisoryAffected versions, risk and source history
Related advisories and sources
Publishers link these advisories through the identifiers shown below. A shared identifier or date alone does not tell you whether your software is affected or whether the advisories describe the same cause.
CVE-2026-18844 · CVE overview and exploitation status
cve-program · CVE-2026-18844
Related changes
Linked by CVE identifierNo related changes are available here.