Mira Hormone Monitor, Mira Android App Missing authentication for critical function
cve-program · CVE-2026-66875
PublishedIn the Mira hormone monitor device firmware v1.7.1.47 build 01070147, a remote unauthenticated attacker within BLE range (approximately 10–30 meters) can silently rebind the device to an attacker-controlled account, extract stored hormone measurements in cleartext, cause a denial-of-service via malformed or undocumented command opcodes, and passively track the user via a static random BLE address that never rotates.
Read the publisher's advisoryAffected versions, risk and source history
Related advisories and sources
Publishers link these advisories through the identifiers shown below. A shared identifier or date alone does not tell you whether your software is affected or whether the advisories describe the same cause.
CVE-2026-66875 · CVE overview and exploitation status
cve-program · CVE-2026-66875
Related changes
Linked by CVE identifierNo related changes are available here.